Troubleshoot the Bus

When instruments stop responding, return unexpected data, or the bus hangs, mcpyvisa provides several diagnostic tools to identify and recover from the problem. This guide covers the most common failure modes and how to resolve them.

flowchart TD
    START["check_srq"]
    ASSERTED{"SRQ<br/>asserted?"}
    FIND["find_rqs<br/>(identify source)"]
    FOUND{"Device<br/>found?"}
    SPOLL["serial_poll<br/>(read status byte)"]
    INTERPRET["Interpret status bits<br/>RQS / ESB / MAV"]
    MANUAL["serial_poll all<br/>known instruments"]
    DONE(["Handle condition"])
    CLEAR(["No service<br/>needed"])

    START --> ASSERTED
    ASSERTED -- "Yes" --> FIND
    ASSERTED -- "No" --> CLEAR
    FIND --> FOUND
    FOUND -- "Yes" --> SPOLL
    FOUND -- "No" --> MANUAL
    SPOLL --> INTERPRET
    MANUAL --> INTERPRET
    INTERPRET --> DONE

    style START fill:#78350f,stroke:#d97706,color:#fde68a
    style ASSERTED fill:#78350f,stroke:#d97706,color:#fde68a
    style FOUND fill:#78350f,stroke:#d97706,color:#fde68a
    style DONE fill:#166534,stroke:#22c55e,color:#dcfce7
    style CLEAR fill:#334155,stroke:#94a3b8,color:#e2e8f0

GPIB instruments signal that they need attention by asserting the SRQ (Service Request) line. This is a shared bus signal — any instrument can assert it.

> Check if any instrument is requesting service on bench-a

The LLM calls check_srq("bench-a"). If SRQ is asserted, mcpyvisa automatically calls find_rqs to identify which device raised it:

SRQ ASSERTED on bench-a -- device at GPIB0::5::INSTR requesting service (status=0x50)

If find_rqs cannot pinpoint the source:

SRQ ASSERTED on bench-a -- use serial_poll to identify the source

Serial polling reads the status byte from an instrument without disrupting its operation. This is the primary way to determine what an instrument needs.

> Serial poll the ammeter

The LLM calls serial_poll("ammeter"):

Serial poll:
ammeter (GPIB0::5::INSTR): status=0x50 (80) [SRQ]
> Serial poll everything on bench-a

The LLM calls serial_poll("bench-a") with no specific instrument, which polls all instruments discovered on that backend:

Serial poll on bench-a:
GPIB0::1::INSTR (voltmeter): status=0x00 (0)
GPIB0::5::INSTR (ammeter): status=0x50 (80) [SRQ]
GPIB0::22::INSTR (psu): status=0x00 (0)

The status byte is an 8-bit value where each bit has a defined meaning per IEEE 488.2:

BitMaskMeaning
70x80Instrument-specific
60x40RQS — device is requesting service (SRQ source)
50x20ESB — event status bit (check *ESR? for details)
40x10MAV — message available in output buffer
3—00x0FInstrument-specific

A status byte of 0x50 means bits 6 and 4 are set: the instrument is requesting service (RQS) and has a message available (MAV).

flowchart TD
    START["Instrument not<br/>responding"]
    SCAN{"discover_instruments<br/>-- found?"}
    HW["Check cables,<br/>power, address"]
    ERR["Query SYST:ERR?"]
    ERRQ{"Errors<br/>found?"}
    CLS["Send *CLS<br/>(clear status)"]
    RST["Send *RST<br/>(reset)"]
    RETEST{"Responding<br/>now?"}
    OK(["Recovered"])
    ESCALATE(["Hardware issue --<br/>check physical layer"])

    START --> SCAN
    SCAN -- "No" --> HW
    SCAN -- "Yes" --> ERR
    HW --> ESCALATE
    ERR --> ERRQ
    ERRQ -- "Yes" --> CLS
    ERRQ -- "No" --> RST
    CLS --> RETEST
    RST --> RETEST
    RETEST -- "Yes" --> OK
    RETEST -- "No" --> ESCALATE

    style START fill:#7f1d1d,stroke:#ef4444,color:#fecaca
    style OK fill:#166534,stroke:#22c55e,color:#dcfce7
    style ESCALATE fill:#7f1d1d,stroke:#ef4444,color:#fecaca
    style SCAN fill:#78350f,stroke:#d97706,color:#fde68a
    style ERRQ fill:#78350f,stroke:#d97706,color:#fde68a
    style RETEST fill:#78350f,stroke:#d97706,color:#fde68a
  1. Verify the instrument is on the bus

    > Discover instruments on bench-a without identification

    If the instrument does not appear, the problem is physical: check the GPIB cable, make sure the instrument is powered on, and verify its GPIB address setting.

  2. Check the instrument’s error queue

    > Query SYST:ERR? on the DMM

    Read errors until you get 0,"No error". Some instruments lock up when their error queue is full.

  3. Clear the instrument’s status

    > Send *CLS to the DMM

    The LLM calls instrument_write("dmm", "*CLS"). This clears status registers and error queues.

  4. Reset the instrument

    > Reset the DMM

    The LLM calls instrument_reset("dmm"), sending *RST. Wait several seconds for the instrument to reconfigure.

A timeout means the instrument did not respond within the configured read_timeout_ms. Common causes:

CauseSolution
Instrument is busy with a long measurementIncrease read_timeout_ms before the query
Wrong command — instrument has nothing to sendUse instrument_write instead of instrument_query
Instrument is in local modeSend instrument_remote to put it back in remote mode
Bus contention from another controllerSend interface_clear to reassert control
ESP32 brownout during rapid commandsIncrease inter_command_delay_ms in the config

When the GPIB bus stops responding to any commands:

  1. Assert Interface Clear

    > Send interface clear on bench-a

    The LLM calls interface_clear("bench-a"). This pulses the IFC line for 150 microseconds, forcing all devices to release the bus and making this backend the Controller-In-Charge.

  2. Send a device clear

    > Clear the DMM

    The LLM calls instrument_clear("dmm"). This sends Selected Device Clear (SDC) to the instrument, which typically aborts any pending operation and clears input/output buffers. To clear all devices on the bus, omit the instrument argument:

    instrument_clear("bench-a")
  3. Re-discover to verify recovery

    > Discover instruments on bench-a

    If instruments reappear, the bus is recovered. If not, the problem is likely hardware (cable, connector, or a misbehaving instrument pulling bus lines low).

These two tools address different layers of the problem:

ToolWhat it doesWhen to use
instrument_clear (with instrument)Sends Selected Device Clear (SDC) to one instrumentAn instrument is stuck or has a full error queue
instrument_clear (backend only)Sends Universal Device Clear (DCL) to all instrumentsMultiple instruments are unresponsive
interface_clearPulses IFC line — resets bus interfaces, not instrument stateBus is completely hung, no commands get through
flowchart LR
    subgraph interface_clear["interface_clear (IFC)"]
        direction TB
        IFC_A["Resets handshake lines"]
        IFC_B["Clears talker/listener<br/>assignments"]
        IFC_C["Reasserts Controller-<br/>In-Charge"]
    end

    subgraph instrument_clear["instrument_clear (DCL / SDC)"]
        direction TB
        BC_A["Clears instrument<br/>input/output buffers"]
        BC_B["Aborts pending<br/>operations"]
        BC_C["Resets instrument<br/>parser state"]
    end

    IFC_SCOPE["Bus interface<br/>layer"]
    BC_SCOPE["Instrument<br/>state"]

    interface_clear -.-> IFC_SCOPE
    instrument_clear -.-> BC_SCOPE

    style IFC_SCOPE fill:#334155,stroke:#94a3b8,color:#e2e8f0
    style BC_SCOPE fill:#334155,stroke:#94a3b8,color:#e2e8f0
    style interface_clear fill:#1e293b,stroke:#d97706,color:#fde68a
    style instrument_clear fill:#1e293b,stroke:#d97706,color:#fde68a

For systematic diagnosis, use the built-in prompt:

> Use the troubleshoot-bus prompt for the ammeter on bench-a

This guides the LLM through a six-step diagnostic sequence:

  1. Verify bus connectivity with discovery
  2. Check instrument identity with *IDN?
  3. Read status byte with serial poll
  4. Clear error state with *CLS and SYST:ERR?
  5. Reset the instrument with *RST
  6. Bus-level recovery with interface_clear and instrument_clear

The prompt exits at the first step that resolves the problem, so minor issues are fixed quickly.

For suspected cable or connector problems, use ar488_diagnostic to write known bit patterns directly to the bus lines:

> Run AR488 diagnostic on bench-a, write 0xAA to the data bus

The LLM calls ar488_diagnostic("bench-a", mode=0, value=0xAA):

Diagnostic: writing 0xAA to data bus on bench-a for 10 seconds

This holds the pattern on the bus for 10 seconds, giving you time to probe individual lines with an oscilloscope or logic analyzer. Mode 0 writes to the data bus (DIO1-DIO8), mode 1 writes to the control bus (ATN, IFC, REN, SRQ, EOI).

The ESP32’s brownout detector is disabled in the AR488 firmware because GPIB bus activity causes transient voltage dips. However, if the ESP32 is powered from a USB port with marginal current capacity, you may see:

  • Backend disconnects mid-sequence
  • Corrupted responses (partial or garbled data)
  • The adapter stops responding and requires a power cycle

Mitigations:

  • Use a powered USB hub or a USB port with reliable 500mA+ capacity
  • Increase inter_command_delay_ms to 20—50 ms to reduce peak current draw
  • For WiFi-connected adapters, use a dedicated 5V power supply rather than USB power
  • Keep GPIB cable runs short (under 4 meters per segment, under 20 meters total)